SERWIS APARATURY MEDYCZNEJ __________________________________BIOMEDICAL EQUIPMENT - REPAIRS - MAINTENANCE

STRONA GŁÓWNA

 

FIZYKOTERAPIA

 

EiE Otwock :

Galvatronic GT-1C

Pulsotronic ST-4M

Diatronic DT-7B

Interdynamic ID-4C

Interdynamic ID-8C

Diatronic DT-10b

Pulsotronic ST-6D

Magnetronic MF-8 

Magnetronic MF-10 

Lasertronic LT-30

 

Famed:

Stymat S-110                                                                                 

Stymat S-120                                                                            

Diadynamik S-200                                                                         

Diadynamik S-210                                                                     

Interdynamik S-300                                                            

Terapuls GS-200                                                                          

Terapuls GS-220                                                                       

 

 

  

      Fizykoterapeutyczne stymulatory prądowe (diadynamik, interdynamik, galwatronik, stymat) powinny być okresowo sprawdzane pod względem prawidłowych wskazań dawkowanych prądów.

Częstym problemem są zużyte elektrody gumowe które zwiększają impedancję połączenia aparatu z ciałem pacjenta co może prowadzić do nieprawidłowej stabilizacji prądu stymulującego (odczuwanie porażenia prądowego przez pacjenta). Zużycie gumowych elektrod nie jest widoczne gołym okiem. W miarę eksploatacji zwiększa się rezystancja czynnej powierzchni elektrody.

Biostymulatory laserowe po kilku latach intensywnej eksploatacji mogą wyrażnie zmniejszać moc wiązki promieniowania. Wynika to ze spadku emisji diody laserowej. Większość aparatów posiada wbudowany tester do sprawdzania poziomu energii.

 
Powered by dzs.pl & Hosting & Serwery